Dräger Infinity Explorer C700
cpe:2.3:h:draeger:infinity_explorer_c700:*:*:*:*:*:*:*, +1 more
A privilege escalation vulnerability has been identified in the Dräger Infinity Explorer C700. This vulnerability allows attackers to exit kiosk mode and access the underlying operating system by interacting with a specific dialog. Exploiting this kiosk escape could enable unauthorized control over the operating system, potentially leading to the device displaying incorrect information or no information at all from the connected Delta Family patient monitor.
Exploitation of this vulnerability could result in unauthorized access to the operating system, allowing for manipulation of system functions and potentially disrupting the accurate display of patient monitoring data.